Elements and Performance Criteria
- Identify the measured quantity and the uncertainty components
- List uncertainty components that are associated with each input in the equation
- Determine the size of each uncertainty component
- Use calibration reports, manufacturer's specifications, quality control and validation data, and experimental data to collect other available information on the uncertainty components
- Reduce each uncertainty component to a standard uncertainty
- Calculate the standard uncertainties
- Calculate an expanded uncertainty to the required confidence level
- Calculate a combined standard uncertainty
- Determine an appropriate coverage factor based on the degrees of freedom associated with each uncertainty component
- Calculate the expanded uncertainty
- Report the expanded uncertainty
- Report the confidence level and coverage factor
- Determine the appropriateness of the size of the expanded uncertainty relative to the tolerance or required accuracy of the test
- Determine the fitness for purpose of the expanded uncertainty relative to the use of the measurement result